• DocumentCode
    2949388
  • Title

    Young´s modulus measurement of the porcine ligamentum flavum for evaluating the deformation during epidural insertion

  • Author

    Naemura, K.

  • Author_Institution
    Tokyo Univ. of Technol., Tokyo, Japan
  • fYear
    2010
  • fDate
    Aug. 31 2010-Sept. 4 2010
  • Firstpage
    3923
  • Lastpage
    3925
  • Abstract
    In order to check validity of the porcine ligamentum flavum as a substitute of human tissue for evaluating the epidural needles, the thickness and the Young´s modulus were measured, and a finite element analysis was carried out. Twenty-four specimens were stained by the Elastica-van Gieson method. The average value of the thickness was 1.2 mm. The tensile tester was employed for measuring the Young´s modulus. A cross sectional area was obtained by X-ray CT image. Young´s modulus was 1.6 MPa. The measured properties of a porcine ligamentum flavum showed a reduced thickness and smaller Young´s modulus than the human ligamentum flavum. On a finite element analysis, the shape of the ligamentum flavum was modeled as a square plate 5 mm in width and 5 mm in depth. The thickness and the Young´s modulus were set at the measured value for a porcine model, and value from the literature for a human model. For simplicity, a constant load of 10 N was applied at the center of each model. The porcine model exhibited deformation 1.7 times that of the human model.
  • Keywords
    Young´s modulus; biological tissues; biomechanics; biomedical measurement; computerised tomography; deformation; diagnostic radiography; finite element analysis; physiological models; Elastica-van Gieson method; X-ray CT image; Young modulus; deformation; epidural insertion; finite element analysis; human tissue; porcine ligamentum flavum; tensile tester; Deformable models; Finite element methods; Humans; Load modeling; Needles; Thickness measurement; Young´s modulus; Animals; Finite Element Analysis; Ligaments; Magnetic Resonance Imaging; Paraffin Embedding; Swine; Tensile Strength; Tomography, X-Ray Computed;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
  • Conference_Location
    Buenos Aires
  • ISSN
    1557-170X
  • Print_ISBN
    978-1-4244-4123-5
  • Type

    conf

  • DOI
    10.1109/IEMBS.2010.5627685
  • Filename