DocumentCode
2965028
Title
Assessment of intracranial pressure
Author
Jensen, Christine ; LaCourse, John
Author_Institution
Biomed. Eng. Center, New Hampshire Univ., Durham, NH, USA
fYear
1988
fDate
10-11 Mar 1988
Firstpage
133
Lastpage
135
Abstract
A feasibility study of a noninvasive intraocular tonometer consisting of a linear voltage differential transformer (LVDT) for measurement of intracranial pressure (ICP) is discussed. The feasibility was assessed by measuring ICP through an artificially created fontanel in pigs. An LVDT deflection corresponds directly with the response of an invasive pressure transducer inserted into the cerebrospinal fluid. The LVDT can also provide a discrete measurement of intracranial pressure. A handheld LVDT tonometer would allow for a quick noninvasive assessment of ICP, and serve as a screening process for infants at risk of abnormal ICP
Keywords
biomechanics; biomedical measurement; pressure measurement; artificially created fontanel; cerebrospinal fluid; infants; intracranial pressure assessment; linear voltage differential transformer; medical measurement; noninvasive intraocular tonometer; pigs; quick noninvasive assessment; screening process; Biomedical measurements; Circuits; Cranial pressure; Detectors; Magnetic cores; Monitoring; Operational amplifiers; Pediatrics; Pressure measurement; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioengineering Conference, 1988., Proceedings of the 1988 Fourteenth Annual Northeast
Conference_Location
Durham, NH
Type
conf
DOI
10.1109/NEBC.1988.19365
Filename
19365
Link To Document