Vocal tract modeling as related to the use of an artificial larynx
Author
Myrick, Reginald ; Yantorno, Robert
Author_Institution
Dept. of Electr. Eng., Temple Univ., Philadelphia, PA, USA
fYear
1993
fDate
18-19 Mar 1993
Firstpage
75
Lastpage
77
Abstract
The placement of an artificial larynx forward in the vocal tract, which results in a very different system from the normal larynx-vocal tract configuration, is considered. The consequences of this excitation relocation are studied. It is found that zeros are introduced into what is otherwise modeled as an all-pole system
Keywords
bioelectric potentials; biomedical electronics; digital simulation; equivalent circuits; frequency response; medical computing; physiological models; poles and zeros; prosthetics; speech synthesis; PSPICE; all-pole system; artificial larynx; electrically driven speech prosthesis; electrolarynx; equivalent network model; excitation relocation; frequency response; larynx-vocal tract configuration; terminal analog model; vocal tract modeling; wave digital filter model; zeros; Acoustic propagation; Cancer; Concatenated codes; Filters; Larynx; Oncological surgery; Prosthetics; Speech analysis; Speech processing; Speech synthesis;