DocumentCode
3061211
Title
Rigid DIFFUSION PHANTOM: Acquisition and simulation
Author
Sakai, Koji ; Azuma, Takashi ; Mori, Susumu
Author_Institution
Center for the Promotion of Excellence in Higher Education, Kyoto University, Japan
fYear
2008
fDate
20-25 Aug. 2008
Firstpage
451
Lastpage
453
Abstract
Diffusion tensor imaging has already been a popular imaging modality for clinical researches. However, to setup an imaging protocol is not straightforward. This problem is critical issue for multi-center studies. We developed anisotropy phantoms and examined the feasibility of using water-filled arrays of channeled silicon plates. Various channel sizes were tested to optimize the diffusion and anisotropy properties. We succeeded in producing a phantom with high anisotropy (FA = 0.8). However, a reason of high axial diffusivity needs further investigations.
Keywords
Anisotropic magnetoresistance; Biomedical imaging; Diffusion tensor imaging; Imaging phantoms; Laminates; Protocols; Rubber; Silicon; Tensile stress; Testing; Computer Simulation; Computer-Aided Design; Equipment Design; Equipment Failure Analysis; Feasibility Studies; Humans; Image Interpretation, Computer-Assisted; Magnetic Resonance Imaging; Models, Theoretical; Phantoms, Imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2008. EMBS 2008. 30th Annual International Conference of the IEEE
Conference_Location
Vancouver, BC
ISSN
1557-170X
Print_ISBN
978-1-4244-1814-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2008.4649187
Filename
4649187
Link To Document