• DocumentCode
    3097684
  • Title

    Silicon substrate ringing in microfabricated ultrasonic transducers

  • Author

    Ladabaum, Igal ; Wagner, Paul ; Zanelli, Claudio ; Mould, John ; Reynolds, Paul ; Wojcik, Greg

  • Author_Institution
    Sensant Corp., San Jose, CA, USA
  • Volume
    1
  • fYear
    2000
  • fDate
    36800
  • Firstpage
    943
  • Abstract
    Experimental and theoretical evidence of silicon substrate ringing in microfabricated ultrasonic transducers is presented. This ringing is clearly observed in immersion transducers with a 650 μm thick substrate at 7 MHz and harmonics. An analytical model of the ringing is introduced, and simulations based on the model are shown to agree with experimental observation. Experimental results are further compared to simulations carried out in time-domain, large-scale PZFlex models and qualitative agreement is demonstrated. The insights gained from the simulations and experiments are used to design and fabricate a device whose ringing mode is eliminated with a backing layer
  • Keywords
    elemental semiconductors; harmonics; micromachining; silicon; substrates; time-domain analysis; ultrasonic transducers; 650 mum; 7 MHz; Si; Si substrate ringing; analytical model; backing layer; design; harmonics; immersion transducers; large-scale PZFlex models; microfabricated ultrasonic transducers; model; ringing mode; simulations; time-domain; Analytical models; Circuit simulation; Electrodes; Equivalent circuits; Frequency; Large-scale systems; Resonance; Silicon; Time domain analysis; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2000 IEEE
  • Conference_Location
    San Juan
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-6365-5
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2000.922696
  • Filename
    922696