  • Abstract
    Traditionally dielectric charging has been viewed as an undesirable feature of Capacitive Micromachined Ultrasound Transducers (CMUTs) because charging can alter performance of devices and lead to permanent collapse. We introduce a beneficial aspect of dielectric charging in CMUTs. We demonstrate CMUTs that may be pre-charged with a shift in collapse voltage so that at zero applied bias voltage they experience an electrostatic deflection due to trapped dielectric charge. We demonstrate efficient pre-collapse operation at zero applied bias voltage with exceptional long-term reliability. Similar existing strategies for pre-charging CMUTs were not implemented for diagnostic frequencies for immersion or tissue-coupled applications.
  • Keywords
    bioMEMS; biomedical transducers; biomedical ultrasonics; electrostatics; micromachining; capacitive micromachined ultrasound transducers; collapse voltage; dielectric charge trapping; electrostatic deflection; low-bias voltage operation; medical applications; precharged CMUT; Biomedical equipment; Dielectrics; Fabrication; Medical services; Testing; Transducers; Voltage measurement; Combine capacitive micromachined ultrasound transducers; MEMS; dielectric charging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2013 IEEE International
  • Conference_Location
    Prague
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4673-5684-8
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2013.0440
  • Filename
    6725118