    EPSAC Predictive Control of Blood Glucose Level in Type I Diabetic Patients

  • Author

    Ionescu, Clara ; De Keyser, Robin

  • Author_Institution
    Electrical energy, Systems and Automation Department at Ghent University, Technologiepark 913, B9052, Gent, Belgium (phone: 32-9-2645608; fax: 32-9-2645603; e-mail: clara@autoctrl.UGent.be).
  • fYear
    2005
  • fDate
    12-15 Dec. 2005
  • Firstpage
    4845
  • Lastpage
    4850
  • Abstract
    An in-house predictive control algorithm has been implemented to control blood glucose level in type I diabetic patients, by controlling the insulin infusion rate to a mechanical pump. The role of the disturbance filter in model-based predictive control is underlined and its possibility to improve control performance is exploited. For comparison purposes, a classic PID controller has been designed via CAD tools. Controller performance was assessed in terms of its ability to track a normoglycemic setpoint (81mg/dL) from initial states of hypo- and hyperglycemia, as well as in response to a meal disturbance. Unconstrained control gave satisfactory results, within imposed constraints. Both strategies are implemented in discrete manner, thus suitable for future implementation in wearable devices.
  • Keywords
    Blood; Design automation; Diabetes; Filters; Insulin; Prediction algorithms; Predictive control; Predictive models; Sugar; Three-term control;