DocumentCode
3121811
Title
Dosimetric Verification of Collapsed Cone Convolution Algorithm in Heterogeneous Media
Author
Guan, Yu-min ; Zhou, Ling-hong ; Zhen, Xin ; Lu, Wen-ting ; Zhang, Shu-xu
Author_Institution
Sch. of Biomed. Eng., Southern Med. Univ., Guangzhou, China
fYear
2010
fDate
18-20 June 2010
Firstpage
1
Lastpage
4
Abstract
The purpose of this study is to investigate the accuracy of dose predicted by cone convolution algorithm. This work describes an experimental verification of two dose calculation algorithms available on the Philips Pinnacle3 (version 8.0) commercial radiotherapy treatment planning system (TPS): Collapsed cone convolution and Fast convolve. Both experiment and TPS were tested for 6 MV photon beam, the fields were 5cm×5cm and 10cm×10cm respectively. Dosimetry was performed with thermoluminescent dosimeters. At the end of this article, we analyzed the difference between calculation and experiment.
Keywords
convolution; dosimetry; radiation therapy; Philips Pinnacle3; collapsed cone convolution algorithm; dosimetric verification; fast convolve; heterogeneous media; photon beam; radiotherapy treatment planning system; Accuracy; Biomedical applications of radiation; Biomedical measurements; Convolution; Electromagnetic scattering; Electron beams; Lungs; Monte Carlo methods; Neoplasms; Nonhomogeneous media;
fLanguage
English
Publisher
ieee
Conference_Titel
Bioinformatics and Biomedical Engineering (iCBBE), 2010 4th International Conference on
Conference_Location
Chengdu
ISSN
2151-7614
Print_ISBN
978-1-4244-4712-1
Electronic_ISBN
2151-7614
Type
conf
DOI
10.1109/ICBBE.2010.5516473
Filename
5516473
Link To Document