• DocumentCode
    3121811
  • Title

    Dosimetric Verification of Collapsed Cone Convolution Algorithm in Heterogeneous Media

  • Author

    Guan, Yu-min ; Zhou, Ling-hong ; Zhen, Xin ; Lu, Wen-ting ; Zhang, Shu-xu

  • Author_Institution
    Sch. of Biomed. Eng., Southern Med. Univ., Guangzhou, China
  • fYear
    2010
  • fDate
    18-20 June 2010
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    The purpose of this study is to investigate the accuracy of dose predicted by cone convolution algorithm. This work describes an experimental verification of two dose calculation algorithms available on the Philips Pinnacle3 (version 8.0) commercial radiotherapy treatment planning system (TPS): Collapsed cone convolution and Fast convolve. Both experiment and TPS were tested for 6 MV photon beam, the fields were 5cm×5cm and 10cm×10cm respectively. Dosimetry was performed with thermoluminescent dosimeters. At the end of this article, we analyzed the difference between calculation and experiment.
  • Keywords
    convolution; dosimetry; radiation therapy; Philips Pinnacle3; collapsed cone convolution algorithm; dosimetric verification; fast convolve; heterogeneous media; photon beam; radiotherapy treatment planning system; Accuracy; Biomedical applications of radiation; Biomedical measurements; Convolution; Electromagnetic scattering; Electron beams; Lungs; Monte Carlo methods; Neoplasms; Nonhomogeneous media;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioinformatics and Biomedical Engineering (iCBBE), 2010 4th International Conference on
  • Conference_Location
    Chengdu
  • ISSN
    2151-7614
  • Print_ISBN
    978-1-4244-4712-1
  • Electronic_ISBN
    2151-7614
  • Type

    conf

  • DOI
    10.1109/ICBBE.2010.5516473
  • Filename