• DocumentCode
    3150331
  • Title

    Verification for EMI test of cardiac pacemaker by portable telephones with an anatomically based human model

  • Author

    Wang, Jianqing ; Ohshima, Takeshi ; Takahashi, Eri ; Fujiwara, Osamu

  • Author_Institution
    Nagoya Inst. of Technol.
  • Volume
    1
  • fYear
    2003
  • fDate
    16-16 May 2003
  • Firstpage
    104
  • Abstract
    Due to the difficulty of electromagnetic interference (EMI) measurement in an actual pacemaker use situation, the recommendations for the management of health risks of implanted pacemaker users from portable telephones have been developed, based mainly on an in-vitro measuring system with a cuboid liquid phantom simulating a homogeneous human body. The validity of such a highly simplified human body model, however, has never been carefully examined. In this paper, with the application of our previously proposed numerical method for the EMI evaluation of pacemakers by portable telephones, we first confirmed Irnich´s finding that the maximum interference distance increases with the third root of the antenna transmitting power for a homogeneous cuboid torso model in order to show the validity of our modeling. We then investigated whether the finding holds true for an actual human body using an anatomically based human model, and confirmed the usefulness of the simple cuboid model in the EMI evaluation for cardiac pacemakers
  • Keywords
    antennas; cardiology; electromagnetic compatibility; electromagnetic interference; finite difference time-domain analysis; pacemakers; phantoms; physiological models; EMI evaluation; EMI testing; FDTD method; anatomically based human model; antenna transmitting power; cardiac pacemaker; cuboid liquid phantom; electromagnetic interference measurement; health risk management; homogeneous cuboid torso model; homogeneous human body; in-vitro measuring system; interference distance; pacemaker implant; portable telephones; Biological system modeling; Electromagnetic interference; Electromagnetic measurements; Humans; Imaging phantoms; In vitro; Pacemakers; Risk management; Telephony; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electromagnetic Compatibility, 2003. EMC '03. 2003 IEEE International Symposium on
  • Conference_Location
    Istanbul
  • Print_ISBN
    0-7803-7779-6
  • Type

    conf

  • DOI
    10.1109/ICSMC2.2003.1428205
  • Filename
    1428205