  • Abstract
    Methods are reviewed for achieving high data rates in bio-implantable devices. Particular attention is given to cortical stimulator arrays for restoring vision. Error-correction codes (ECC) are shown to be essential to obtain reliable operation in next-generation high-rate implants. A survey of reported ECC implementations is presented, and power-vs-performance tradeoffs are revealed after re-scaling to remove differences in technology and clock speed. Recommendations are offered for realizing high-rate ECC circuits within the tight power constraints of implantable device applications.
  • Keywords
    clocks; error correction; prosthetics; bioimplantable devices; bioimplantable electronics; clock speed; cortical stimulator arrays; error correction circuits; error-correction codes; implantable device applications; power constraints; Circuits and systems; Decoding; Error correction codes; Implants; Parity check codes; Receivers; Wireless communication;