• DocumentCode
    3265786
  • Title

    A GTO speed-up inverter for fast-scan magnetic resonance imaging

  • Author

    Mueller, O.M. ; Roemer, P.B. ; Park, J.N. ; Souza, S.P. ; Peters, R. L St

  • Author_Institution
    General Electric Co., Schenectady, NY, USA
  • fYear
    1992
  • fDate
    23-27 Feb 1992
  • Firstpage
    479
  • Lastpage
    486
  • Abstract
    Magnetic resonance imaging (MRI) machines require unique high-power electronic systems for the generation of pulsed B gradient fields, especially for echo planar MRI, where images are taken in 25-30 ms instead of minutes. The speedup inverter described has been implemented with gate turn-off thyristors (GTOs). A novel nonresonant circuit combines the advantages of high-efficiency switch-mode power conversion techniques for the generation of megawatt pulses. An order-of-magnitude increase in available gradient power has been achieved with excellent fidelity and stability
  • Keywords
    biomedical NMR; biomedical electronics; invertors; pulsed power technology; switched mode power supplies; thyristor applications; 25 to 30 ms; GTO; HF; SMPS; biomedical NMR; echo planar MRI; magnetic resonance imaging; nonresonant circuit; power electronics; pulsed B gradient fields; pulsed power technology; speedup inverter; stability; Coils; Inverters; Magnetic resonance imaging; Power generation; Pulse amplifiers; Pulse generation; Pulse power systems; Switches; Thyristors; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Applied Power Electronics Conference and Exposition, 1992. APEC '92. Conference Proceedings 1992., Seventh Annual
  • Conference_Location
    Boston, MA
  • Print_ISBN
    0-7803-0485-3
  • Type

    conf

  • DOI
    10.1109/APEC.1992.228371
  • Filename
    228371