Abstract
A machine built for ultrasound velocity in vivo measurements in order to trace heel bone density changes with outstandingly high precision, even surpassing results one gets from first class radiological methods, is described. Due to the high precision, a special region of interest (ROI) with a very sharply defined sound axis which can be used as a reference against which to observe minute bone mineralization changes has been found. In vivo studies of sound velocities in the heels of normal persons demonstrate that reproducibilities of ±0.1% of equivalent heel bone densities should be possible
Keywords
bioacoustics; biomechanics; biomedical equipment; biomedical ultrasonics; bone; ultrasonic velocity; extremely sensitive ultrasound scanning system; human heel bones; minute mineralisation changes; noninvasive in vivo detection; osteoporosis; region of interest; trace heel bone density changes; ultrasound velocity in vivo measurements; Acoustic transducers; Biological tissues; Bones; Delay; Density measurement; Humans; In vivo; Mineralization; Osteoporosis; Ultrasonic imaging;