DocumentCode
3296671
Title
The effect of scintillator manipulation on a thick crystal gamma detector
Author
Joung, Jinhun ; Malmin, Ron ; Engdahl, John C.
Author_Institution
Siemens Med. Solutions., Peoria, IL
Volume
5
fYear
2005
fDate
23-29 Oct. 2005
Firstpage
2468
Lastpage
2473
Abstract
When a NaI(Tl) based gamma camera is used for high energy single photon imaging or positron imaging, a relatively thick scintillation crystal is required to provide enough stopping power for impinging gamma rays. However, it causes worsening of the image quality due to the wider scintillated light spread and parallax error. To address the image quality degradation problem with a thick crystal, we investigated the effects of crystal manipulation on detector performance. The methods include drilling holes, cutting grooves and making pyramid shapes on the gamma ray entrance surface of the crystal. It has been proven throughout this investigation that the proposed method can improve intrinsic spatial/energy resolution, DOI (depth of interaction) decodeability and uniformity. This is done by proper controlling of the optical characteristics of scintillated light photons. Among the tested methods, entrance side groove cutting found to be the most promising method in terms of improving detector performance and providing cost effective manufacturability as well
Keywords
gamma-ray detection; positron emission tomography; single photon emission computed tomography; solid scintillation detectors; NaI(Tl) based gamma camera; cost effective manufacturability; crystal manipulation; depth-of-interaction decodeability; depth-of-interaction uniformity; detector performance; entrance side groove cutting; high energy single photon imaging; image quality degradation; intrinsic energy resolution; intrinsic spatial resolution; positron imaging; scintillated light photons; scintillator manipulation; thick crystal gamma detector; Cameras; Gamma ray detection; Gamma ray detectors; Gamma rays; Image quality; Optical control; Optical imaging; Photonic crystals; Positrons; Solid scintillation detectors;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2005 IEEE
Conference_Location
Fajardo
ISSN
1095-7863
Print_ISBN
0-7803-9221-3
Type
conf
DOI
10.1109/NSSMIC.2005.1596841
Filename
1596841
Link To Document