    Evaluation of scattering in cone-beam breast computed tomography: A Monte Carlo and experimental phantom study

  • Author

    Mettivier, Giovanni ; Russo, Paolo ; Lanconelli, Nico ; Meo, Sergio Lo

  • Author_Institution
    Dipt. di Sci. Fisiche, Univ. di Napoli Federico II, Naples, Italy
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    2930
  • Lastpage
    2935
  • Abstract
    Monte Carlo simulations of X-ray irradiations of breast phantoms of various sizes as PMMA cylinders of different diameters and a hemi-ellipsoidal PMMA phantom were performed in order to evaluate the 2D distribution of primary and scattered photons and Scatter-to-Primary Ratio (SPR) in projection images in cone-beam breast CT. Irradiation geometry and technique factors were set so as to reproduce the experimental conditions used for validation measurements with a prototype CT breast scanner. In the simulations, performed with GEANT4 software, we varied the phantom diameter and shape, the X-ray tube voltage and added filtration. SPR increases from 0.4 (at 8 cm cylinder diameter) up to 1.5 (14 cm cylinder diameter) at the center of the phantom. In the same phantom, SPR has lower values toward the bases of the cylinder than at its center. The scatter fraction increases by adopting a 50 kVp or higher tube voltages, up to 80 kVp, and by increasing the added filtration. Simulated and measured lateral profiles across a 14 cm cylinder diameter in projection images show a relative deviation of 4%. Simulations show a different distribution of scatter in a 14 diameter cylinder and 14 cm hemi-ellipsoidal phantom, so preventing the use of simple cylindrical geometries when simulating the attenuation of the pendant breast for scatter correction procedures.
  • Keywords
    Monte Carlo methods; X-ray scattering; computerised tomography; phantoms; 2D photon distribution; GEANT4 software; Monte Carlo simulation; PMMA cylinders; Scatter-to-Primary Ratio; X-ray irradiation; X-ray tube voltage; breast phantoms; cone beam breast computed tomography scattering; irradiation geometry; Breast; Computed tomography; Electromagnetic scattering; Geometry; Imaging phantoms; Monte Carlo methods; Particle scattering; Voltage; X-ray imaging; X-ray scattering;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5401611
  • Filename
    5401611