• DocumentCode
    3327443
  • Title

    Unified geometric calibration and image registration for detached small animal SPECT/CT

  • Author

    Zhang, Xuezhu ; Chen, Fangfu ; Li, Yongping ; Wei, Qin ; Zhang, Hualin ; Qi, Yujin

  • Author_Institution
    Shanghai Inst. of Appl. Phys., Grad. Univ. of Chinese Acad. of Sci., Shanghai, China
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    3318
  • Lastpage
    3321
  • Abstract
    The aim of this study is to develop a unified optimization method to estimate the complete parameters about geometric calibration of system misalignment for both cone-beam CT and pinhole SPECT, and parameters about coordinate transformation of image registration for these detached systems. The uniform projection equations are derived for both cone-beam and pinhole imaging geometry. Complete geometric parameters are estimated by point object phantom with a priori relative position information. The cost function is structured as the least-squares about residual error. The implementation of nonlinear estimation utilizes the Powell method so as to constrain the optimization problem of this study to converge to the unique solution. The registration procedure is also implemented by the nonlinear optimization method similar to the geometric calibration, and six parameters are sufficient to register the detached systems for multi-modality image fusion. We designed a general-purpose point object/sources phantom for both geometric calibration and image registration procedures. To validate the method presented, we performed small animal experiment studies based on our Micro-SPECT and Micro-CT systems. The results from the imaging experiments show that the unified estimation method is considerable effective and robust for detached multi-modality Micro-SPECT/CT systems.
  • Keywords
    calibration; computerised tomography; image fusion; image registration; medical image processing; phantoms; single photon emission computed tomography; Powell method; cone beam computerized tomography; coordinate transformation; detached small animal SPECT/CT; geometric calibration; image registration; multimodality image fusion; nonlinear optimization method; pinhole SPECT; residual error; Animals; Calibration; Computed tomography; Cost function; Geometry; Image registration; Imaging phantoms; Nonlinear equations; Optimization methods; Parameter estimation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5401743
  • Filename
    5401743