• DocumentCode
    3329532
  • Title

    Digital Nuclear Magnetic Resonance Acquisition Channel

  • Author

    Baldrighi, Paola ; Castellano, Marco ; Vacchi, Carla ; Canina, Davide ; Golzi, Paolo ; Ferrante, Gianni

  • Author_Institution
    Dept. of Electron., Univ. of Pavia, Pavia
  • fYear
    2008
  • fDate
    3-5 Sept. 2008
  • Firstpage
    399
  • Lastpage
    404
  • Abstract
    A new digital receiver is realized to elaborate signals from nuclear magnetic resonance (NMR). The proposed system replaces as much analog electronics as possible with digital electronics. The receiver can detect NMR signals in a frequency range of 0.5-90 MHz with direct detection and higher frequency with undersampling. The user can set parameters as bandwidth to be observed, gain of receiver and accumulation of results. Signal-to-Noise Ratio of the receiver chain is 60 dB. The system is realized on Field Programmable Gate Array (FPGA) on a Peripheral Component Interconnect (PCI) board. Experimental results obtained are presented.
  • Keywords
    data acquisition; field programmable gate arrays; nuclear instrumentation; nuclear magnetic resonance; peripheral interfaces; digital nuclear magnetic resonance acquisition channel; digital receiver; field programmable gate array; peripheral component interconnect; signal-to-noise ratio; Analog-digital conversion; Field programmable gate arrays; Frequency conversion; Magnetic fields; Magnetic moments; Magnetic noise; Nuclear electronics; Nuclear magnetic resonance; Phase noise; Receivers; Digital NMR System; NMR Signal Detection; undersampling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Digital System Design Architectures, Methods and Tools, 2008. DSD '08. 11th EUROMICRO Conference on
  • Conference_Location
    Parma
  • Print_ISBN
    978-0-7695-3277-6
  • Type

    conf

  • DOI
    10.1109/DSD.2008.103
  • Filename
    4669262