Title
Monte Carlo results from neural networks as an alternative to Compton photons LOR analysis
Author
Michaud, Jean-Baptiste ; Rechka, Sanae ; Brunet, Charles-Antoine ; Lecomte, Roger ; Fontaine, Réjean
Author_Institution
Dept. of Electr. & Comput. Eng., Univ. de Sherbrooke, Sherbrooke, QC, Canada
fYear
2009
fDate
Oct. 24 2009-Nov. 1 2009
Firstpage
3909
Lastpage
3916
Abstract
Sensitivity is critical in small-animal PET, and including more of the discarded detections would increase it. However lowering the energy threshold compromises the spatial resolution. This paper is an update on a method to include triple coincidences in the image without significant image degradation. With the energy threshold set as low as 50 keV, the triple coincidences analyzed are the simplest inter-crystal Compton scatter scenario where one photoelectric 511-keV detection coincides with two detections whose energy sum is also 511-keV. The method uses neural networks instead of traditional Compton interaction mathematical models to compute the proper Line-of-Response (LOR) for that coincidence. The paper revisits the algorithm structure, and in particular the preprocessing steps required to simplify the data fed to the network, preprocessing which improves the LOR computation significantly. The paper then presents Monte-Carlo analysis of the method with various point and cylinder sources. The simulated scanner geometry is purposely made to encompass the very worst-case conditions seen in PET scanners today: small diameter, poor photoelectric fraction, poor 35% FWHM energy resolution. LOR identification error is around 20 to 25% and the sensitivity increase ranges from approximately 70 to 100%. Finally, images were recently obtained, with overall very good quality.
Keywords
Compton effect; Monte Carlo methods; coincidence techniques; medical signal processing; neural nets; positron emission tomography; Compton photon LOR analysis alternative; Monte Carlo analysis; coincidence line of response; electron volt energy 511 keV; energy threshold; intercrystal Compton scatter scenario; neural networks; photoelectric detection; scanner geometry; small animal PET sensitivity; triple coincidences; Computational modeling; Computer networks; Degradation; Electromagnetic scattering; Mathematical model; Monte Carlo methods; Neural networks; Particle scattering; Positron emission tomography; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location
Orlando, FL
ISSN
1095-7863
Print_ISBN
978-1-4244-3961-4
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2009.5401932
Filename
5401932
Link To Document