DocumentCode
3351313
Title
Research on gas diffusion mechanisms of limiting current oxygen sensor at low gas pressure
Author
Jian, Jiawen ; Gao, Jianyuan ; Zhou, Jie
Author_Institution
Coll. of Inf. Sci. & Eng., Ningbo Univ., Ningbo, China
fYear
2010
fDate
26-28 June 2010
Firstpage
3211
Lastpage
3214
Abstract
In this article the gas diffusion mechanisms of the limiting current oxygen sensors was studied. For a limiting current oxygen sensor which the pore diameter d is about 30μm, the limiting current Il was quite linear with ln (1-X02) rather than X02 at 1 atm gas pressure, it showed the normal mechanism was more dominant than the Knudsen one. However at the low gas pressure, the limiting current Il was changed with the variation of total gas pressure P. This reveals the gas diffusion mechanisms is neither simply belonged to the normal diffusion nor Knudsen diffusion at low gas pressure. It may be belonged to a mixed diffusion.
Keywords
Knudsen flow; diffusion; gas sensors; pressure; Knudsen diffusion; gas diffusion; limiting current oxygen sensor; low gas pressure; Boundary conditions; Cathodes; Differential equations; Educational institutions; Gas detectors; Information science; Monitoring; Platinum; Stationary state; limiting current; low gas pressure; oxygen sensor; the normal and Knudsen mechanisms;
fLanguage
English
Publisher
ieee
Conference_Titel
Mechanic Automation and Control Engineering (MACE), 2010 International Conference on
Conference_Location
Wuhan
Print_ISBN
978-1-4244-7737-1
Type
conf
DOI
10.1109/MACE.2010.5535735
Filename
5535735
Link To Document