Abstract
Pulse height spectra were obtained by irradiating 55Fe radioactive X-ray source on an InAs n-i-p APD cooled to 77K. The full width half maximum (FWHM) of the detected photopeak at 5.9 keV improves with increasing reverse bias. The FWHM decreases rapidly with gain from 2 keV at M = 1 to ~950eV at M = 5.2, improving the minimum detectable energy and resolution of the APD. The larger than expected FWHM is due to non-optimised packaging that leads to increased leakage current of the detector.
Keywords
X-ray apparatus; X-ray detection; avalanche photodiodes; indium compounds; ionisation chambers; radioactive sources; FWHM; InAs; Pulse height spectra; X-ray detection; avalanche photodiodes; electron volt energy 2 keV; electron volt energy 5.9 keV; electron volt energy 950 eV; full width half maximum; irradiating 55Fe radioactive X-ray source; minimum detectable energy; n-i-p APD; non-optimised packaging; photopeak; temperature 77 K;