• DocumentCode
    3384814
  • Title

    Study on the application of robust statistical techniques in interlaboratory comparisons for oil emission spectrometers

  • Author

    Pei, Wen ; Tian, Hong-Xiang ; Liu, Yu

  • Author_Institution
    Coll. of Naval Archit. & Power, Naval Univ. of Eng., Wuhan, China
  • Volume
    3
  • fYear
    2010
  • fDate
    9-10 Oct. 2010
  • Firstpage
    19
  • Lastpage
    23
  • Abstract
    In order to insure the reliability and the validity of the detection capability and detection data, 10 oil emission spectrometry laboratories which distribute in different areas tested 3 lubricating oil samples composed by different elements content, and interlaboratory comparisons for the detection capability of oil emission spectrometers were carried out. Based on robust statistical techniques, the detection data was evaluated and the evaluation results of robust Z-score were compared with the results of En. The results shows that the robust Z-score method had 86.67% hit rate, 7.33% failed rate, 6.00% omitted rate; while the En method had 31.33% hit rate, 66.67% failed rate, 2.00% omitted rate. the robust Z-score method is more suitable to be applied to interlaboratory comparisons for oil emission spectrometers than En method.
  • Keywords
    laboratory techniques; lubricating oils; spectrometers; statistical analysis; Z-score method; lubricating oil samples; oil emission spectrometers; robust statistical techniques; Copper; Iron; Lead; Nickel; Silicon; Tin; Zinc; En; interlaboratory comparisons; robust Z-score; spectrometer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Future Information Technology and Management Engineering (FITME), 2010 International Conference on
  • Conference_Location
    Changzhou
  • Print_ISBN
    978-1-4244-9087-5
  • Type

    conf

  • DOI
    10.1109/FITME.2010.5654766
  • Filename
    5654766