• DocumentCode
    3386575
  • Title

    Estimation of body fluid losses during abdominal surgery by impedance measurement technique

  • Author

    Çakmak, Günnur ; Bahadirlar, Yildurm ; Ulgen, Yekta

  • Author_Institution
    Inst. of Biomed. Eng., Bogazici Univ., Istanbul, Turkey
  • fYear
    1992
  • fDate
    18-20 Aug 1992
  • Firstpage
    92
  • Lastpage
    94
  • Abstract
    Electrical impedances of the arm, the leg and the trunk are measured separately from the right side at 50 KHz (1 mA r.m.s.), from patients undergoing abdominal surgery, using the tetrapolar (current drive-voltage sense) measurement technique with disposable ECG electrodes. Experimental results strongly support the clinical usefulness of this technique
  • Keywords
    bioelectric phenomena; biomedical measurement; electric impedance measurement; surgery; 50 KHz; abdominal surgery; body electrical impedance; body fluid losses; disposable ECG electrodes; impedance measurement technique; tetrapolar measurement technique; Abdomen; Biomedical engineering; Biomedical measurements; Electrical resistance measurement; Electrocardiography; Electrodes; Extracellular; Frequency; Impedance measurement; Surgery;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering Days, 1992., Proceedings of the 1992 International
  • Conference_Location
    Istanbul
  • Print_ISBN
    0-7803-0743-7
  • Type

    conf

  • DOI
    10.1109/IBED.1992.247087
  • Filename
    247087