DocumentCode
3393130
Title
Performance of the automatic detection of local activation times on unipolar cardiac electrograms in man
Author
Massé, Stéphane ; Savard, Pierre ; Shenasa, Mohammad ; Cardinal, René ; Pagé, Pierre
Author_Institution
Inst. de Genie Biomed., Hopital Sacre-Coeur, Montreal, Que., Canada
fYear
1988
fDate
4-7 Nov. 1988
Abstract
The performance of a simple algorithm for the detection of local activation times on cardiac electrograms was evaluated. Unipolar epicardial and endocardial electrograms were recorded in 21 patients undergoing arrhythmia surgery. Local activation times were manually edited and used as standard references. The quasioptimal performance showed a sensitivity and a positive prediction accuracy of 89%.<>
Keywords
computerised pattern recognition; electrocardiography; medical computing; algorithm; arrhythmia surgery; endocardial electrograms; epicardial electrograms; man; pattern recognition; sensitivity; standard references; unipolar cardiac electrograms;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location
New Orleans, LA, USA
Print_ISBN
0-7803-0785-2
Type
conf
DOI
10.1109/IEMBS.1988.94431
Filename
94431