• DocumentCode
    3407583
  • Title

    Prevention of left atrial wall suction in a continuous-flow total artificial heart by controlling right pump output

  • Author

    Olegario, P.S. ; Yoshizawa, M. ; Tanaka, A. ; Abe, K. ; Yambe, T. ; Nitta, S. ; Takeda, H.

  • Author_Institution
    Tohoku Univ., Sendai, Japan
  • Volume
    5
  • fYear
    2002
  • fDate
    5-7 Aug. 2002
  • Firstpage
    2855
  • Abstract
    One of the problems in continuous-flow total artificial heart control is atrial wall suction. In this study we developed a control algorithm which uses the right pump to prevent left atrial wall suction from occurring. Results of in vitro and in vivo tests indicate that prevention of left atrial wall suction is possible using this method and that it can be performed simultaneously with physiological control of the artificial heart.
  • Keywords
    artificial organs; cardiology; haemodynamics; medical control systems; artificial hearts control; atrial suction; automatic control; continuous-flow; in vitro tests; in vivo tests; physiological control; right pump; total artificial heart; Algorithm design and analysis; Animal structures; Artificial heart; Automatic control; Blood; Control systems; Energy efficiency; Humans; Pressure control; Radio control;
  • fLanguage
    English