Abstract
In this paper the behaviour of some devices commonly used for current measurement is examined by means of a number of experimental tests, considering either sinusoidal or distorted conditions. Traditional current transformers, transducers based on the Hall effect and resistive shunts combined with isolation amplifiers have been considered. The comparison is performed by taking into account amplitude errors and time shifts, as well as deviations caused by the different harmonic contents between input and output quantities. The presented investigation aims to find out the main characteristics that make each transducer suitable or unsuitable in different practical situations
Keywords
Hall effect transducers; current transformers; electric current measurement; measurement errors; power system measurement; Hall effect transducers; LV systems; active current transducers; amplitude errors; comparative analysis; current measurement devices; current transformers; distorted conditions; harmonic content deviations; isolation amplifiers; passive current transducers; power systems; resistive shunts; sinusoidal conditions; time shifts; Current measurement; Hall effect; Instruments; Optical amplifiers; Performance analysis; Power system harmonics; Power system measurements; Testing; Transducers; Voltage;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 1999. IMTC/99. Proceedings of the 16th IEEE
Conference_Location
Venice
ISSN
1091-5281
Print_ISBN
0-7803-5276-9
Type
conf
DOI
10.1109/IMTC.1999.777010