DocumentCode
3445651
Title
Fast Synchronous Ac-Dc-Transfer with Thermal Converters
Author
Mohns ; Kiirten, W.G.
Author_Institution
Ihlenfeld Phys.-Technische Bundesanstalt - PTB, Braunschweig
fYear
2004
fDate
38139
Firstpage
190
Lastpage
191
Abstract
The most accurate measurements of the rms (root mean square) values of ac (alternating current) voltages are performed by means of thermal converters (TC) (Klonz, 2002) and digital sampling procedures (Ihlenfeld, 2001 and Ihlenfeld, et. al., 2003) (at low frequencies) with uncertainties of the order of some fractions of muV/V. The main drawback of using TCs relies on the long measuring times required, of a few minutes in most cases. In the following, a simple method for reducing the measuring time and to substantially speed up ac-dc transfer measurements will be described and documented with results of a comparison measurement
Keywords
AC-AC power convertors; measurement standards; measurement uncertainty; voltage measurement; alternating current voltage; digital sampling; fast synchronous AC-DC transfer; root mean square voltage; thermal converters; Analog-digital conversion; Cutoff frequency; Digital-to-frequency converters; Frequency measurement; Sampling methods; Signal processing; Switches; Time measurement; Velocity measurement; Zero voltage switching;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements Digest, 2004 Conference on
Conference_Location
London
Print_ISBN
0-7803-8494-6
Electronic_ISBN
0-7803-8494-6
Type
conf
DOI
10.1109/CPEM.2004.305525
Filename
4097180
Link To Document