• DocumentCode
    3447487
  • Title

    Influence of variable propagation speed on high-definition compound imaging

  • Author

    Herment, Alain ; Guglielmi, Jean-Paul ; Peronneau, Pierre

  • Author_Institution
    INSERM, Paris, France
  • fYear
    1988
  • fDate
    4-7 Nov. 1988
  • Firstpage
    1124
  • Abstract
    The principle of a high-definition ultrasound imaging based on a compound acquisition of B-scans with a sectorial echograph is presented. Data processing is based on both deconvolution and reflection tomography. The images obtained by using this method, with respect to those of two-dimensional deconvolution, show the following: (1) an improvement in definition close to 9; (2) a gain in signal-to-noise ratio of about 4; and (3) a decrease in computations of 12. The effects of a limited acquisition window and of variable propagation speed are discussed.<>
  • Keywords
    biomedical ultrasonics; computerised picture processing; medical diagnostic computing; B scan compound acquisition; data processing; deconvolution; high-definition compound imaging; limited acquisition window; reflection tomography; sectorial echograph; signal-to-noise ratio; ultrasound imaging; variable propagation speed;