Title
An improved multicrystal 2-D BGO detector for PET
Author
Rogers, J.G. ; Taylor, A.J. ; Rahimi, M.F. ; Nutt, R. ; Andreaco, M. ; Williams, C.W.
Author_Institution
TRIUMF, Vancouver, BC, Canada
fYear
1991
fDate
2-9 Nov. 1991
Firstpage
1609
Abstract
The authors evaluate and compare two 2-D array detectors for PET (positron emission tomography). Both consist of an 8*8 array of small BGO crystals coupled to a 2*2 array of photomultiplier tubes. The depth of the crystals is 3 cm in one detector and 2 cm in the other. The 2-cm detector is obviously superior in terms of material costs, but it is also superior in energy resolution per crystal, and in its ability to identify clearly the crystal containing the primary interaction. The authors present a flexible and robust algorithm for crystal identification in such array detectors. The prospect of obtaining still better spatial resolution from such block detectors, with increased numbers of crystals, is discussed.<>
Keywords
biomedical equipment; computerised tomography; radioisotope scanning and imaging; 2 cm; 3 cm; block detectors; crystal identification; energy resolution; flexible robust algorithm; material costs; medical diagnostic imaging; medical instrumentation; multicrystal 2D BGO detector; nuclear medicine; photomultiplier tubes; positron emission tomography; Costs; Crystalline materials; Crystals; Detectors; Energy resolution; Photomultipliers; Positron emission tomography; Robustness; Sensor arrays; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
Conference_Location
Santa Fe, NM, USA
Print_ISBN
0-7803-0513-2
Type
conf
DOI
10.1109/NSSMIC.1991.259186
Filename
259186
Link To Document