• DocumentCode
    3461774
  • Title

    The status of the ICR/RAL, BaF2-TMAE positron camera

  • Author

    Wells, K. ; Ott, R.J. ; Suckling, J. ; Bateman, J.E. ; Stephenson, R. ; Connolly, J.F.

  • Author_Institution
    Inst. of Cancer Res., R. Marsden Hospital, Belmont, UK
  • fYear
    1991
  • fDate
    2-9 Nov. 1991
  • Firstpage
    1668
  • Abstract
    A novel BaF/sub 2/-TMAE (tetrakis dimethylaminoethylene) detector has been developed which can simultaneously achieve 5-ns timing, 5-mm spatial resolution, and 20% detection efficiency. The chamber can operate in a stable condition for periods of several hours. Results of projection imaging are also presented. The projection images obtained with the 100-cm/sup 2/ sensitive area detector represent the first occasion a distributed source has been used in conjunction with a BaF/sub 2/-TMAE detector for positron imaging.<>
  • Keywords
    biomedical equipment; cameras; computerised tomography; radioisotope scanning and imaging; ICR/RAL, BaF/sub 2/-TMAE positron camera; detection efficiency; detector; distributed source; medical diagnostic imaging; medical instrumentation; nuclear medicine; positon emission tomography; projection imaging; spatial resolution; tetrakis dimethylaminoethylene; Cameras; Circuits; Crystals; Detectors; Electrons; Insulation; Positrons; Spatial resolution; Vacuum systems; Wire;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference, 1991., Conference Record of the 1991 IEEE
  • Conference_Location
    Santa Fe, NM, USA
  • Print_ISBN
    0-7803-0513-2
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1991.259198
  • Filename
    259198