    The blood pressure measurement device with convex micro sensor array for reducing a cross talk

  • Author

    Endo, Y. ; Dohi, T.

  • Author_Institution
    Chuo Univ., Tokyo, Japan
  • fYear
    2015
  • fDate
    21-25 June 2015
  • Firstpage
    1719
  • Lastpage
    1722
  • Abstract
    In this study, we fabricate a the blood pressure measurement device based on the arterial tonometry method with a convex micro pressure sensor array for reducing a cross talk of the sensor. The size of our device is 30 × 30 × 5.8 mm3. It has 11 pressure sensors that are 130 × 250 × 20 μm3 in size. Because each pressure sensor element has a convex surface shape, the pressure measured by each sensor element has little effect on the others. Thus, a clear blood pressure pulse wave can be measured by reducing the cross talk of the sensor.
  • Keywords
    biomedical equipment; biomedical measurement; blood pressure measurement; microsensors; pressure sensors; sensor arrays; arterial tonometry; blood pressure measurement device; blood pressure pulse wave; convex micropressure sensor array; cross talk; pressure sensor element; Arrays; Blood pressure; Pressure measurement; Pressure sensors; Sensitivity; Surface waves; Convex micro sensor array; arterial tonometry method; blood pressure measurement; cross talk;