Establishment of 40,000 A Rogowski coil calibration system and its uncertainty analysis
Author
Kim, Yoon Hyoung ; Jung, Jae Kap ; Kang, Jeon Hong ; Lee, Sang Hwa ; Han, Sang Ok
Author_Institution
Chungnam Nat. Univ., Daejeon, South Korea
fYear
2010
fDate
13-18 June 2010
Firstpage
686
Lastpage
687
Abstract
System for calibrating Rogowski coil(RC) up to primary current of 40,000 A have been established by employing standard current transformer(CT), CT comparator and voltage to current converter(VCC). The expanded uncertainty is 0.021 % ~ 0.230 % for ratio error and 0.3 min ~ 1.9 min for phase displacement in the primary current ranges of 10 ~ 40,000 A.
Keywords
calibration; convertors; current transformers; Rogowski coil calibration system; current 10 A to 40000 A; current transformer comparator; standard current transformer; time 0.3 min to 1.9 min; uncertainty analysis; voltage to current converter; Calibration; Coils; Electromagnetic measurements; Uncertainty;