DocumentCode
351831
Title
A technique for measuring scintillator total electron energy resolution
Author
Mengesha, W. ; Valentine, J.D.
Author_Institution
Nucl. & Radiol. Eng. Program, Georgia Inst. of Technol., Atlanta, GA, USA
Volume
1
fYear
1999
fDate
1999
Firstpage
162
Abstract
A technique for determining the total electron energy resolution (ηe,tot) of scintillators is described. The technique is based on measurement of the full-width at half maximum of the scintillator coincidence energy spectrum using the previously developed Compton coincidence technique (CCT). For determining ηe,tot , the CCT was modified to accommodate variable collimation of the scattered photons. MCNP simulations were used to guide development of data analysis techniques. Preliminary measurements indicate ηe,tot variation as the inverse square root of the electron energy over a limited energy range
Keywords
Monte Carlo methods; coincidence techniques; scintillation counters; ηe,tot; Compton coincidence technique; FWHM; MCNP simulation; full-width at half maximum; scintillator; total electron energy resolution; Collimators; Current measurement; Detectors; Electrons; Energy measurement; Energy resolution; Geometry; Pulse measurements; Scattering; Solids;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium, 1999. Conference Record. 1999 IEEE
Conference_Location
Seattle, WA
ISSN
1082-3654
Print_ISBN
0-7803-5696-9
Type
conf
DOI
10.1109/NSSMIC.1999.842468
Filename
842468
Link To Document