• DocumentCode
    3532581
  • Title

    Quantitative investigation of inter-crystal scatter and penetration in the GE Discovery RX PET/CT scanner using Monte Carlo simulations

  • Author

    Zeraatkar, Navid ; Ay, Mohammad R. ; Sarkar, Saeed ; Geramifar, Parham ; Rahmim, Arman

  • Author_Institution
    Res. Center for Sci. & Technol. in Med., Tehran Univ. of Med. Sci., Tehran, Iran
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    2403
  • Lastpage
    2408
  • Abstract
    As the dimensions of crystal elements in modern PET systems is getting smaller, the fraction of events undergoing inter-crystal scattering (ICS) and penetration increases. This has deteriorative impact on the spatial resolution of images. However, with the utilization of statistical image reconstruction methods in PET, compensation for ICS and penetration impacts is feasible by accurately modeling their effects in the projection space. We aim to perform such a work for the GE Discovery RX (DRX) PET/CT system. In this work, we investigated ICS and penetration in the DRX for acquiring a quantitative view of their respective contributions. We analyzed the events in the form of coincidences instead of single photons. Also as a novelty, we discriminated between the origins of the event mispositiong. For this, we applied the GATE (Geant4 Application for Tomographic Emission) Monte Carlo toolkit and used our previously validated model of DRX. Numerous points in different positions were analyzed inside the field of view (FOV) of DRX. Finally, using geometrical symmetries of the DRX and interpolation, the fraction of ICS and penetration could be determined for all possible positions inside the FOV. The results revealed that the notable variations in quantitative behavior of ICS and penetration occurred with varying radial positions; the fraction of ICS/penetration-induced mispositioned coincidences out of true coincidences ranged from 28.7% at the center to 57.8% at the edge of the transaxial FOV. This comprehensive quantification of ICS and penetration not only provides a deeper understanding of their respective contributions, but is also aimed to be utilized in refining the system matrix in the image reconstruction task to achieve resolution modeling in the scanner.
  • Keywords
    Monte Carlo methods; image reconstruction; image resolution; medical image processing; positron emission tomography; statistical analysis; GATE Monte Carlo toolkit; GE Discovery RX PET-CT scanner; Geant4 application; Monte Carlo simulations; intercrystal penetration; intercrystal scatter; spatial resolution; statistical image reconstruction methods; tomographic emission; Biomedical imaging; Crystals; Detectors; Image reconstruction; Integrated circuit modeling; Photonics; Positron emission tomography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5874218
  • Filename
    5874218