  • Abstract
    A breath-hold (BH) acquisition method is easy of operation and recently studied actively. In the method, a patient is asked to hold his/her breath for 10 to 30 s as the image acquisition is performed. This is repeated several times and the obtained images are summed. In practice, however, a patient cannot necessarily hold his/her breath at the same timing of breathing. In such a case, the summed images still has a blur. In this paper, we propose to apply our image registration method developed before for respiratory-gated images to BH images. The BH images are nonlinearly motion-corrected so as to match a reference BH image and summed. Through the experiment with ten patient BH image data, we confirmed that the proposed method is effective in obtaining a non-blurred image.
  • Keywords
    biomechanics; deformation; image registration; medical image processing; pneumodynamics; positron emission tomography; breath-hold PET imaging; breath-hold acquisition method; breathing; deformation; image acquisition; image registration method; nonblurred imaging; nonlinearly motion-correction; respiratory-gated imaging; Computed tomography; Image registration; Lungs; Pixel; Positron emission tomography; Timing; Tumors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5874482
  • Filename
    5874482