DocumentCode
3548521
Title
Evaluation of rotating slant hole SPECT mammography with respect to planar scintimammography using Monte Carlo simulation methods
Author
Liu, Chi ; Tsui, Benjamin M W ; Baird, William H. ; Xu, Jingyan ; Wang, Yuchuan ; Frey, Eric C.
Author_Institution
Dept. of Radiol., Johns Hopkins Univ.
Volume
7
fYear
2004
fDate
16-22 Oct. 2004
Firstpage
4063
Lastpage
4067
Abstract
The purpose of this study is to evaluate mammoSPECT using a rotating 4-segment slant-hole (R4SSH) collimator compared to planar scintimammography (SM). Four pairs of R4SSH and parallel-hole collimators with the same spatial resolutions of 0.7 cm, 0.9 cm, 1.1 cm and 1.3 cm at 16.5 cm, 16.8 cm, 17.0 cm and 17.1 cm respectively, which are the distances to the centers of the common volume-of-view (CVOV) of the R4SSH collimators, were used. Monte Carlo simulated noise-free and noisy planar SM and R4SSH projection images of the left breast were obtained from the 4-D NCAT phantom. The breast was placed within the CVOV of the R4SSH collimator and 14 cm from the LEHR collimator. A 0.78 cm diameter lesion with a lesion/background activity ratio of 9:1 was placed in the lateral, central and medial region of the breast. The R4SSH projection data were reconstructed using a 3D iterative OSEM method. Image quality was evaluated in terms of lesion contrast and FWHM, background noise level and contrast-to-noisy ratio (CNR). The results indicated that, for each collimator pair, the lesion contrast increases and FWHM of lesion decreases in the R4SSH SPECT images while the noise level increases with iteration number. When compared to the planar SM images obtained with equal acquisition time, the R4SSH SPECT images have higher noise level and poorer spatial resolution but significantly higher lesion contrast especially for the medial lesions. Using appropriate iteration number, R4SSH SPECT images can provide higher CNR for medial and even central lesions, but lower CNR for lateral lesions. We conclude that RMSSH SPECT can potentially provide improved image quality as compared to planar SM
Keywords
Monte Carlo methods; biological organs; collimators; image reconstruction; image resolution; iterative methods; mammography; medical image processing; phantoms; single photon emission computed tomography; 3D iterative OSEM method; 4-D NCAT phantom; FWHM; Monte Carlo simulation methods; R4SSH projection images; acquisition time; background noise level; common volume-of-view; contrast-to-noisy ratio; image quality; iteration number; left breast; medial lesions; noise-free planar SM projection images; noisy planar SM projection images; parallel-hole collimators; planar scintimammography; rotating 4-segment slant hole SPECT mammography; spatial resolutions; Breast; Collimators; Image quality; Imaging phantoms; Lesions; Mammography; Monte Carlo methods; Noise level; Samarium; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2004 IEEE
Conference_Location
Rome
ISSN
1082-3654
Print_ISBN
0-7803-8700-7
Electronic_ISBN
1082-3654
Type
conf
DOI
10.1109/NSSMIC.2004.1466786
Filename
1466786
Link To Document