DocumentCode
3567398
Title
Complex microwave noise parameter measurement using a coaxial radiometer
Author
Sinclair, M.W. ; Williams, G.L.
Author_Institution
NPL, DRA, Malvern, UK
fYear
1994
fDate
7/15/1894 12:00:00 AM
Firstpage
42430
Lastpage
42433
Abstract
A technique has been described for the measurement of the complex noise parameters of an active device using a computer controlled coaxial radiometer designed for noise source calibration. The uncertainties on the measured parameters are fully traceable to National Standards. The current values yield an uncertainty of +or-0.2 dB in noise figure at the 95% confidence level. This is limited largely by the uncertainty on the primary noise standard to which all the noise source calibrations are traceable. This primary standard will be replaced in the near future and a lower uncertainty level is predicted. The system used is limited to the frequency band 10 MHz to 18 GHz, but the technique can in principle be applied to any total power radiometer.<>
Keywords
calibration; computerised instrumentation; electric noise measurement; measurement standards; microwave measurement; radiometers; 10 MHz to 18 GHz; National Standards; active device; complex microwave noise parameter measurement; computer controlled coaxial radiometer; noise source calibrations; primary noise standard; total power radiometer; Calibration; Measurement standards; Microwave measurements; Noise measurement; Radiometry;
fLanguage
English
Publisher
iet
Conference_Titel
What's New in Microwave Measurements (Digest No.1994/042), IEE Colloquium on
Type