• DocumentCode
    3567487
  • Title

    Using a hybrid pixel detector for dynamic radiography

  • Author

    Pedersen, Frands ; Sandborg, M.

  • Author_Institution
    Dept. of Radiat. Sci., Uppsala Univ., Sweden
  • Volume
    2
  • fYear
    1996
  • Firstpage
    1239
  • Abstract
    The hybrid pixel detector has recently been introduced in digital radiography. A Monte Carlo simulation was performed with the purpose of comparing the photon counting and energy integration registration strategies, using the detective quantum efficiency DQE. The results were compared with the input screen of an image intensifier. A cascaded model of the hybrid pixel detector is outlined
  • Keywords
    Monte Carlo methods; diagnostic radiography; image intensifiers; medical diagnostic computing; photon counting; position sensitive particle detectors; Monte Carlo simulation; detective quantum efficiency; digital radiography; dynamic radiography; energy integration registration; hybrid pixel detector; image intensifier; input screen; photon counting; Displays; Frequency; Image intensifiers; Radiation detectors; Radiography; Sensor arrays; Spatial resolution; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium, 1996. Conference Record., 1996 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-3534-1
  • Type

    conf

  • DOI
    10.1109/NSSMIC.1996.591634
  • Filename