Many button type and strip line type beam position monitors are employed in the South Hall Ring (SHR) at MIT-Bates. It is desired to calibrate and routinely check the readouts of those monitors by simulating the beam. A wire setup was built to serve this purpose. It consists of a button type beam position monitor body with a wire running through it. To eliminate end-effects both ends are extended with 10 inch long uniform beam pipes. RF matching is carefully performed with ECCOSORB AN 75 microwave absorber forms. The wire is positioned 3 mm offset with respect to the X-pair buttons and symmetric with respect to the Y-pair buttons. Thus the same setup can be used to check both sensitivity and zero-offset
Keywords
calibration; detector circuits; electron accelerators; nuclear electronics; particle beam diagnostics; storage rings; 10 in; 3 mm; ECCOSORB AN 75 microwave absorber forms; MIT-Bates; RF matching; South Hall Ring; Y-pair buttons; beam position monitors; button type beam position monitors; calibration; sensitivity; strip line type beam position monitors; uniform beam pipes; wire setup calibration; zero-offset; Calibration; Electronic circuits; Fixtures; Linear accelerators; Particle beams; Radio frequency; Stripline; Strips; Structural beams; Wire;
fLanguage
English
Publisher
ieee
Conference_Titel
Particle Accelerator Conference, 1995., Proceedings of the 1995
Print_ISBN
0-7803-2934-1
Type
conf
DOI
10.1109/PAC.1995.505609
Filename
505609
Link To Document