• DocumentCode
    3601425
  • Title

    Novel, Wirelessly Controlled, and Adaptive Artificial Urinary Sphincter

  • Author

    Hached, Sami ; Saadaoui, Zineb ; Loutochin, Oleg ; Garon, Andre ; Corcos, Jacques ; Sawan, Mohamad

  • Author_Institution
    Electr. Eng. Dept., Polytech. Montreal, Montreal, QC, Canada
  • Volume
    20
  • Issue
    6
  • fYear
    2015
  • Firstpage
    3040
  • Lastpage
    3052
  • Abstract
    Severe stress urinary incontinence is a nonnegligible complication that affects men as well as women. When initial therapies are not sufficient, implantation of artificial urinary sphincters (AUS) is considered the gold standard treatment. This hydromechanical device applies constant pressure on the urethra to keep urine inside the bladder. Although recovery from incontinence is full or partial, this approach has certain disadvantages. In addition to the difficulties and discomfort encountered when operating the device, insufficient pressure, urethral atrophy, and erosion from constant pressure on the urethra may necessitate revision surgery. In this study, we present a novel, remote-controlled AUS with variable urethral pressure that improves continence and avoids revision surgery. The proposed design eliminates the manual pump, facilitating implantation in men and women. It is compatible with current AUS. The device requires very little power and its parameters can be monitored and updated post-operatively. It has been tested on pig bladders ex vivo. Different pressure control algorithms have been developed, and expected performances have been achieved. Design challenges and experimentation results are reported.
  • Keywords
    artificial organs; medical control systems; pressure control; surgery; telecontrol; AUS; adaptive artificial urinary sphincter; different pressure control algorithm; hydromechanical device; manual pump; pig bladders; remote-controlled AUS; revision surgery; urethral atrophy; urine; variable urethral pressure; Batteries; Implants; Microcontrollers; Power demand; Switches; Transceivers; Wireless communication; AMS 800; artificial urinary sphincter; implantable electrical medical device; stress urinary incontinence; wireless control;
  • fLanguage
    English
  • Journal_Title
    Mechatronics, IEEE/ASME Transactions on
  • Publisher
    ieee
  • ISSN
    1083-4435
  • Type

    jour

  • DOI
    10.1109/TMECH.2015.2389254
  • Filename
    7047837