• DocumentCode
    3631859
  • Title

    Analysis of Magnetic Resonance signal in inhomogenous main magnetic field

  • Author

    V. Emre Arpinar;B. Murat Eyuboglu

  • Author_Institution
    M?hendislik Fak?ltesi, Elektrik ve Elektronik M?hendisli?i B?l?m?, Orta Do?u Teknik ?niversitesi, Turkey
  • fYear
    2009
  • fDate
    5/1/2009 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Nuclear magnetic resonance (NMR) systems with inhomogenous main magnetic fields, which are used to analyze material properties, have been developed before. Therefore, magnetic resonance imaging (MRI) systems with inhomogenous main magnetic fields can be possibly developed to image tissues. To investigate the possibility of MRI in inhomogenous main magnetic field, magnetic resonance (MR) signal has to be analyzed. In this study, MR signal´s parameters deviation due to inhomogeneity level change in the main magnetic field of a MRI system is investigated. Analyzed MR signal parameters are signal peak, energy and relaxation constant. Deviations of these parameters, because of inhomogeneity level change in the main magnetic field, are related to change in MRI system´s properties. Thus, starting from the MRI system´s properties, tolerable inhomogeneity level in main magnetic field can be found.
  • Keywords
    "Signal analysis","Magnetic analysis","Magnetic resonance","Magnetic fields","Magnetic resonance imaging","Nuclear magnetic resonance","Magnetic properties","Material properties","Radio frequency"
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering Meeting, 2009. BIYOMUT 2009. 14th National
  • Print_ISBN
    978-1-4244-3605-7
  • Type

    conf

  • DOI
    10.1109/BIYOMUT.2009.5130278
  • Filename
    5130278