  • Title

    A comparative biomechanical analysis of Pretzel: A novel arthroscopic suture knot

  • Author

    A.N. Turkoglu;U. Akgun;F. Ates;C.A. Yucesoy;M. Karahan

  • Author_Institution
    Biyomedikal M?hendisli?i Enstit?s?, Bo?azi?i ?niversitesi, 34342 Bebek, ?stanbul, T?rkiye
  • fYear
    2009
  • fDate
    5/1/2009 12:00:00 AM
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    The purpose of this paper is to evaluate the biomechanical properties of a novel arthroscopic knot, Pretzel knot to be used in shoulder surgeries and compare it with two knots taken as standard: Samsung Medical Center(SMC) and Nicky´s knots. Two groups have been formed: unbacked group (Pretzel and SMC knots), backed group (Pretzel, SMC and Nicky´s knots backed with 3 RHAPs). Both groups have been tested with load to failure test and cyclic loading test. Parameters considered for both groups were for loop security, mechanism of failure, permanent loosening, load at failure and elongation at failure. There was significant improvement in loop security for Pretzel compared to SMC knot in unbacked case. Also elongation at failure in backed case is significantly lower for Pretzel compared to SMC whereas all other results were similar. Results showed that Pretzel knot may be safely used in arthroscopic surgeries and shows an improvement in loop security which gives an advantage in surgeries on pretensioned tissues.
  • Keywords
    "Testing","Sliding mode control","Surges","Security","Shoulder","Medical tests"
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering Meeting, 2009. BIYOMUT 2009. 14th National
  • Print_ISBN
    978-1-4244-3605-7
  • Type

    conf

  • DOI
    10.1109/BIYOMUT.2009.5130281
  • Filename
    5130281