• DocumentCode
    3635067
  • Title

    Intracoronary Doppler versus new method of the assessment of coronary flow using computer analysis of conventional coronary angiograms

  • Author

    B. Gorecka;Z. Dzielinska;J. Jodkowski;W. Ruzyllo;P. Szmaus;M. Dabrowski

  • Author_Institution
    Nat. Inst. of Cardiology, Warsaw, Poland
  • fYear
    1996
  • Firstpage
    405
  • Lastpage
    408
  • Abstract
    The aim of the study was validation of a mean brightness method and new coronary flow index CFI for the assessment of coronary flow by means of intracoronary Doppler measurements. This method utilizes the dynamics of coronary angiograms and is based on the computer analysis of conventional coronary angiograms. The contractual measure of blood flow is percent change of mean brightness of coronary artery in unit of time: cardiac cycle (%/cycle) or time in seconds (%/sec). Coronary flow was investigated in 10 patients with angiographically normal coronary artery (NCA group) and in 17 patients with single vessel stenosis (SVS group). In the SVS group, measurements were repeated twice: before and after PTCA procedure. In both groups statistically significant, positive correlations were found between both methods. In conclusion, the new CFI is very reliable, simple and easy to obtain parameters able to assess coronary flow and can be very useful in the case of comparative studies.
  • Keywords
    "Arteries","Brightness","Cardiology","Angiography","Doppler measurements","Blood flow","Time measurement","Wire","Transducers","Fluid flow measurement"
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 1996
  • ISSN
    0276-6547
  • Print_ISBN
    0-7803-3710-7
  • Type

    conf

  • DOI
    10.1109/CIC.1996.542559
  • Filename
    542559