• DocumentCode
    3638639
  • Title

    Real time conductance catheter system based on FPGA

  • Author

    María de los Á. Gómez López;Juan M. Olivera;Myriam C. Herrera

  • Author_Institution
    Electrical, Electronic and Computational Engineering Department, National University of Tucumá
  • fYear
    2010
  • Firstpage
    2525
  • Lastpage
    2528
  • Abstract
    This paper presents a digital conductance catheter system based on FPGA (Field Programmable Gate Arrays) to measure real time ventricular volumes for experimental use. The system performance is realized in three stages: a) conductance measurements using a well-known set of resistance connected to each catheter section; b) volumes measurements in glass containers of well-known geometry that contain a solution with a resistivity similar to the blood, and c) comparison between the previous analog conductance catheter system and this novel FPGA based system.
  • Keywords
    "Catheters","Volume measurement","Field programmable gate arrays","Containers","Glass","Voltage measurement","Calibration"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2010 Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Print_ISBN
    978-1-4244-4123-5
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/IEMBS.2010.5626874
  • Filename
    5626874