• DocumentCode
    3650500
  • Title

    Comparative analysis of materials during design process of prosthetic foot by using plastic injection molding simulation FE tool

  • Author

    M. V. Candal;O. Pelliccioni;M. Bermeo;C. Müller-Karger

  • Author_Institution
    Universidad Simó
  • fYear
    2013
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    A comparative analysis between polymeric materials suggested for a new design of a Dynamic Energy Return (DER) prosthetic foot is presented. This component, developed in the R+D Biomechanics Group of the Universidad Simón Bolívar, conforms a solid block of some 400 cm3, which is unusual in injected plastic parts and requires particular requirements not covered by conventional standards for correct design of an injection mold. Numerical tests were conducted using a plastic injection molding simulation tools. Results related to Filling Time of the cavity mold, Cycle Time, Volumetric Shrinkage, Warpage and Sink Marks of molded part in Polyoxymethylene (POM), polyamide (PA) and polypropylene (PP) are presented and discussed contrasted with the recommendations of the technical bibliography reports, usually for injection of thin-walled parts.
  • Keywords
    "Materials","Conferences","Couplings","IEEE catalog","Prosthetics","Foot"
  • Publisher
    ieee
  • Conference_Titel
    Health Care Exchanges (PAHCE), 2013 Pan American
  • ISSN
    2327-8161
  • Print_ISBN
    978-1-4673-6254-2
  • Electronic_ISBN
    2327-817X
  • Type

    conf

  • DOI
    10.1109/PAHCE.2013.6568274
  • Filename
    6568274