• DocumentCode
    3684242
  • Title

    MRI of biopsy needles by susceptibility mapping based on Wiener filter F and L1-regularization

  • Author

    Tao Yang;Jim Ji

  • Author_Institution
    Department of Electrical and Computer Engineering, Texas A&
  • fYear
    2015
  • Firstpage
    1576
  • Lastpage
    1579
  • Abstract
    Biopsy needles are devices that have been used for intravenous therapy. However, the high susceptibility of needles results in signal loss and distortion which makes the location of needles hard to identify in the MRI images. A variety of approaches has been proposed to quantify the susceptibility of the materials being imaged because susceptibility is an intrinsic property that can be used to make a good contrast between different materials. Although previous techniques on susceptibility mapping seem quite effective, they usually consume too much time due to the iterations. In this paper, a method based on Wiener filter is developed to improve the speed of susceptibility mapping. Simulation and phantom experiment demonstrate the effectiveness and high efficiency of the proposed method.
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7318674
  • Filename
    7318674