DocumentCode
3684527
Title
Measuring cervical vertebra movements using kinect sensor
Author
Xiaolong Ma;Guanghua Xu;Min Li;Jun Xie;Longting Chen;Wei Pei
Author_Institution
School of Mechanical Engineering, Xi´an Jiaotong University Xi´an 710049 China
fYear
2015
Firstpage
2771
Lastpage
2774
Abstract
The incidence of cervical vertebra spondylosis is increasing with the pace of life while there is no high-efficient measurement method. This study proposed a non-contact, high-efficient, and high-precisionmethod measuring the cervical vertebra activities which obtains the cervical vertebra activities by combining the function of skeletal tracking and face tracking. A comparison study between the proposed and traditional methods has been conducted here. Statistics shows that the efficiency of the proposed measurement method at least is 6 times faster than modified square shaped goniometers that is the represent of traditional measurement, and it´s angular error in average is 0.35°. Moreover, the proposed measuring method has been tested in hospital clinical environment and the results show that the proposed method can be used easily and have little interference.
Keywords
"Motion measurement","Face","Measurement uncertainty","Shape measurement","Joints","Rotation measurement","Eyebrows"