DocumentCode
3684657
Title
Effect of intra-aortic balloon pump on coronary blood flow during different balloon cycles support∶ A computer study
Author
Thin Pa Pa Aye;Zwe Lin Htet;Thamvarit Singhavilai;Phornphop Naiyanetr
Author_Institution
Department of Biomedical Engineering, Faculty of Engineering, Mahidol University, Thailand
fYear
2015
Firstpage
3303
Lastpage
3306
Abstract
Intra-aortic balloon pump (IABP) has been used in clinical treatment as a mechanical circulatory support device for patients with heart failure. A computer model is used to study the effect on coronary blood flow (CBF) with different balloon cycles under both normal and pathological conditions. The model of cardiovascular and IABP is developed by using MATLAB SIMULINK. The effect on coronary blood flow has been studied under both normal and pathological conditions using different balloon cycles (balloon off; 1:4; 1:2; 1:1). A pathological heart is implemented by reducing the left ventricular contractility. The result of this study shows that the rate of balloon cycles is related to the level of coronary blood flow.
Keywords
"Heart","Mathematical model","Blood flow","Computational modeling","Pathology"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7319098
Filename
7319098