DocumentCode
3685238
Title
Accuracy of heart rate variability estimation by photoplethysmography using an smartphone: Processing optimization and fiducial point selection
Author
V. Ferrer-Mileo;F. Guede-Fernandez;M. Fernández-Chimeno;J. Ramos-Castro;M. A. García-González
Author_Institution
Group of Biomedical and Electronic Instrumentation, Department of Electronic Engineering, Universitat Politè
fYear
2015
Firstpage
5700
Lastpage
5703
Abstract
This work compares several fiducial points to detect the arrival of a new pulse in a photoplethysmographic signal using the built-in camera of smartphones or a photoplethysmograph. Also, an optimization process for the signal preprocessing stage has been done. Finally we characterize the error produced when we use the best cutoff frequencies and fiducial point for smartphones and photopletysmograph and compare if the error of smartphones can be reasonably be explained by variations in pulse transit time. The results have revealed that the peak of the first derivative and the minimum of the second derivative of the pulse wave have the lowest error. Moreover, for these points, high pass filtering the signal between 0.1 to 0.8 Hz and low pass around 2.7 Hz or 3.5 Hz are the best cutoff frequencies. Finally, the error in smartphones is slightly higher than in a photoplethysmograph.
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7319686
Filename
7319686
Link To Document