DocumentCode
3685627
Title
Quantification of velocity reduction after flow diverter placement in intracranial aneurysm: An ex vivo study with 3D printed replicas
Author
Jeff R. Anderson;Richard Klucznik;Orlando Diaz;Y. Jonathan Zhang;Gavin W. Britz;Robert G. Grossman;Christof Karmonik
Author_Institution
MRI Core Facilities at the Houston Methodist Research Institute, TX 77030 USA
fYear
2015
Firstpage
7300
Lastpage
7303
Abstract
Phase contrast MRI (pcMRI) was used to measure flow before and after placement of a flow diverter (n = 3). Decreases from 18% to 31% in flow velocity were seen in the inflow jet of the aneurysms. Flow patterns were also compared. It was observed that the gross aneurysmal flow patterns were maintained after flow diverter placement despite decreased fluid velocities. All measurements were carried out in 3D printed aneurysm replicas.
Keywords
"Aneurysm","Magnetic resonance imaging","Computational fluid dynamics","Three-dimensional displays","Yttrium","Neck","Neurosurgery"
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
ISSN
1094-687X
Electronic_ISBN
1558-4615
Type
conf
DOI
10.1109/EMBC.2015.7320077
Filename
7320077
Link To Document