• DocumentCode
    3685627
  • Title

    Quantification of velocity reduction after flow diverter placement in intracranial aneurysm: An ex vivo study with 3D printed replicas

  • Author

    Jeff R. Anderson;Richard Klucznik;Orlando Diaz;Y. Jonathan Zhang;Gavin W. Britz;Robert G. Grossman;Christof Karmonik

  • Author_Institution
    MRI Core Facilities at the Houston Methodist Research Institute, TX 77030 USA
  • fYear
    2015
  • Firstpage
    7300
  • Lastpage
    7303
  • Abstract
    Phase contrast MRI (pcMRI) was used to measure flow before and after placement of a flow diverter (n = 3). Decreases from 18% to 31% in flow velocity were seen in the inflow jet of the aneurysms. Flow patterns were also compared. It was observed that the gross aneurysmal flow patterns were maintained after flow diverter placement despite decreased fluid velocities. All measurements were carried out in 3D printed aneurysm replicas.
  • Keywords
    "Aneurysm","Magnetic resonance imaging","Computational fluid dynamics","Three-dimensional displays","Yttrium","Neck","Neurosurgery"
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society (EMBC), 2015 37th Annual International Conference of the IEEE
  • ISSN
    1094-687X
  • Electronic_ISBN
    1558-4615
  • Type

    conf

  • DOI
    10.1109/EMBC.2015.7320077
  • Filename
    7320077