• DocumentCode
    3687893
  • Title

    Beam shaping systems for small animal proton therapy

  • Author

    Lena Mouawad;Marc Rousseau;David Brasse;Ziad El Bitar;Ahmad Osman;Mohamad Khalil;Lena Mouawad;Ziad Francis

  • Author_Institution
    Institut Pluridisciplinaire Hubert Curien, Strasbourg University, CNRS, Strasbourg, France
  • fYear
    2015
  • Firstpage
    174
  • Lastpage
    177
  • Abstract
    Pre-clinical research using small animal models is of crucial importance in cancer research. Small animal radiotherapy platforms have been established allowing the evaluation of therapies and treatment plans in the pre-clinical setting before applying them in human radiotherapy. This work is done in the aim of developing a proton therapy treatment plan for small animals. It deals in particular with the optimization, using Monte Carlo simulations, of a beam shaping device that conforms the dose profile to a subcutaneous tumor five millimeters deep. Three models are compared: A range compensator, attenuator slabs each with a specific collimator, and attenuator slabs with a fixed-size collimator. Using Aluminium slabs of different thicknesses, that can be inserted manually one by one before the tumor in a way to cover all the tumor volume, with a collimator to shield unwanted protons, is judged to be the best method considering the available equipment and the simple mechanical feasibility of such a system.
  • Keywords
    "Tumors","Collimators","Animals","Attenuators","Protons","Slabs","Radiation effects"
  • Publisher
    ieee
  • Conference_Titel
    Advances in Biomedical Engineering (ICABME), 2015 International Conference on
  • ISSN
    2377-5688
  • Electronic_ISBN
    2377-5696
  • Type

    conf

  • DOI
    10.1109/ICABME.2015.7323280
  • Filename
    7323280