• DocumentCode
    3692255
  • Title

    Real-time ophthalmic imaging with a handheld, 20-MHz annular array

  • Author

    Jeffrey A. Ketterling;Daniel Gross;Ronald H. Silverman

  • Author_Institution
    Lizzi Center for Biomedical Engineering, Riverside Research, New York, United States of America
  • fYear
    2015
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    Ophthalmic ultrasound is almost exclusively performed using single-element, handheld transducers. Linear-array systems are not feasible for ophthalmology for the foreseeable future. Annular arrays permit linear-array image quality but without the system complexity and cost. Here, an annular-array transducer previously demonstrated using a cumbersome immersion-scan technique (i.e., water bath coupled to eye) was integrated into a handheld probe that permitted contact scanning of the human eye. A custom, 5-element, 20-MHz annular array based on a 25-μm, P(VDF-TrFE) membrane was packaged in a stainless steel shell and mounted in a commercial, handheld, mechanical probe specifically designed for clinical ophthalmic imaging. The array had a focal length of 25.8 mm and an aperture of 9 mm. The array was characterized using a 25-μm wire and a calibrated hydrophone. After establishing the acoustic parameters were within safety limits, human subjects were imaged. Human-subject scanning was performed by placing the tip of the fluid-filled probe in contact with the sclera.
  • Keywords
    "Probes","Arrays","Acoustics","Imaging","Wires","Transducers","Real-time systems"
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2015 IEEE International
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2015.0254
  • Filename
    7329245